CE-6004n-60V300A-H
Specification
Test System Information
| Parameter | Data |
|---|---|
| Channels | 4 |
| Parallel Connection | Support up to 4 channels in parallel |
| Input Power Supply | 3PH-AC380V ±15%, 50/60 ±5Hz |
| Power Factor | ≥ 99% (full load) |
| THDi | ≤ 5% (full load) |
| Input Power | 84.7 kW |
| Input Current | 128.7 A / single |
| Overall System Efficiency (Max) | 90% |
| Noise | ≤ 65 dB |
| Size (W×D×H) | 950 × 800 × 1950 mm |
| Weight | ≈ 444 kg |
| Power Control Module Type | MOSFET |
| Input Power Wiring | Three-phase five-wire system |
| Input Protection | Anti-surge, anti-islanding, over/under-frequency & voltage, open-phase, etc. |
| Ingress Protection | IP20 |
Measurement & Output
| Parameter | Data |
|---|---|
| Voltage/Current Sampling | Four-wire (same port for charge & discharge) |
| Voltage Range | Charge 0–60 V; Discharge 3–60 V; Min discharge 3 V |
| Voltage Accuracy / Resolution | ±0.02% of FS; 24-bit |
| Current Ranges | 60 A / 120 A / 240 A / 300 A |
| Current Accuracy | ±0.05% of FS (independent range) |
| CV Cut-off Current | 0.05% of FS (per independent range) |
| Current Resolution | 24-bit |
| Single-Channel Output Power | 18 kW |
| Total Output Power | 72 kW |
| Current Response Time (10%FS→90%FS) | ≤ 3 ms |
| Current Conversion Time (−90%FS→90%FS) | ≤ 6 ms |
| Minimum Step Time | 0.1 s |
| Input Impedance | ≥ 1 MΩ |
Modes & Functions
| Parameter | Data |
|---|---|
| Charge/Discharge Modes | CCC/CCD, CVC/CVD, CC-CVC/CC-CVD, CPC/CPD, CP-CVC/CP-CVD, CV-CRD, CRD, VSS, CSS |
| Cut-off Conditions | Voltage, Current, ΔTime, Capacity, −ΔV |
| Simulation | Current, Power; continuous charge/discharge switching; cut-off: Time, step line; steps: up to 1,000,000 |
| Pulse Step | Current/Power; min pulse 50 ms; up to 32 pulses; charge/discharge switch; cut-off: Voltage, ΔTime |
| DCIR | By calculation |
| Safety Protection | Software: set safety limits (V/I min-max, delay…); Hardware: anti-reverse, OVP/OCP/OTP, etc. |
Data & Analysis
| Parameter | Data |
|---|---|
| Step Setting Method | Form editing |
| Recording Frequency | 100 Hz (10 Hz when two AUX channels share one main channel) |
| Database | MySQL |
| Data Output | Excel, TXT (curve templates; customization supported) |
| Loop Test Range | 1–65535 |
| Steps per Loop | ≤ 255 |
| Loop Nesting | ≤ 10 |
Communication
| Parameter | Data |
|---|---|
| Host Communication | TCP/IP protocol |
| Interface | Ethernet |
| Lower Computer Bandwidth | 1 M |
| Host Communication Baud | 10–100 M (adaptive) |
| Networking | LAN via switches/routers |
| Communication Expansion (Optional) | CAN, RS485, BMS (DBC); supports integration with environmental chamber, water chiller, pressure fixture |
AUX (Optional)
| Parameter | Data |
|---|---|
| Temperature Channel | Type-T: −70 to 260 °C; accuracy ±1 °C; resolution 0.1 °C |
| Voltage Channel | 0–5 V; accuracy ±0.05% of FS; resolution 0.1 mV |
| Usage | Bind with main V/I data; measured temperature can be used for control/protection |
Environmental Requirements
| Parameter | Data |
|---|---|
| Operating Temperature | −10 to 40 °C (accuracy guaranteed within 25 ± 5 °C; drift 0.005% of FS/°C) |
| Storage Temperature | −20 to 50 °C |
| Operating Humidity | ≤ 70% RH (no condensation) |
| Storage Humidity | ≤ 80% RH (no condensation) |
| Altitude | < 2000 m; above 2000 m, derate: every +100 m → −1 °C operating temp |