You can export the testing result, set the “Export Type”, “File Format”, “File Name”, and “path”. There are two ways to do it: Click the Excel icon in the menu bar. Right-click the data display area, and select “Export”.
Double-click the yellow bounding rectangle at the bottom, you can set what cycle range you would like to display.
Real-time data can be checked during test process!! Select the channel you want to check; right click, simply choose view data. If you have more questions, please contact me with: collin@newarebattery.com
Display all the current data of working step in each channel, and record the last data in per working step. Right-click the table, and select the contents that you want to copy. You can copy the data to the Excel or other documents. “Step View Setting”
Grading the capacities of batteries according to different aspects of the battery performance. Grading setting: 1.”Grading mode”: “discharge”. Grading the batteries according to the data of discharging working step. Note: the function of “all” and “charge” are temporarily not available. 2.”Cycle”: “last cycle”. Grading the batteries based on the data of the last cycle. Note: the functions of “whole” and
Battery capacity test is one of functions for Neware battery testing equipment. All of models can support this function include BTS 3000,BTS 4000,BTS 5000, BTS 8000, BTS 9000, IGBT….. Sample test: Battery model and specification: Li-ion 18650, 2000mA, 3.7V How dose the battery capacity test work? 1 Fully recharge battery until it reached it’s rated voltage(normally 4.2V); Set up cut-off
“Channel transfer” can transfer working steps and data from one channel to another one. The aim of channel transfer is to avoid single channel to be terminated accidentally caused by physical damage。 1 Select 2 channels(one testing channel and one unused channel) 2 Right click, click “Move”,then the channel been transferred will indicate “Stopped”. If you have more questions, please
Select Server IP and middle machine model first, and then click the “History Data”. 2. On the “History Data” interface, select “Research”, and then we can see all information of this middle machine of this Server. 3. All history data of this middle machine will be ordered based on the starting time, select the serial number, the interface on the
Select the channel to be operated, right-click and select “Single Start”. On the “start” interface, we can set the working step, record condition, and protect condition. Click OK, the channel will start working. *Note: The “start” interface contains “Standard Settings” and “Professional Settings”. “Standard Settings” : set the working step as a whole. “Professional Settings” : every single working step