You can customize the curves/plots you want in our BTSDA software. I’ ll give an example of the method about getting Time- current & voltage curve for you. You can get other kind of curves/plots by the same way. Following screenshot is a typical NDA(Neware data analyzer) file opened in our software BTSDA. And let’s use following steps to create/customize
How to get dQdV curve/plot when you tested cells on Neware cyclers? And how to make the curve smoother? Many Neware cycler users asked this question. BTSDA is the free software from Neware which is used to view the test result, it can help you to understand the battery performance better, and it can export data to txt/xls/csv, what’s more,
Users can set up parameter based on their own requirements, including setting unit, setting log light, cycle layer, step layer, record layer, setting background color. To set up the parameter, simply right click. Setting unit Setting log light Cycle layer Step layer Record layer Setting background color If you have more questions, please contact me with: collin@newarebattery.com
User can copy all the data from software based on their requirements. If you have more questions, please contact me with: collin@newarebattery.com
This video show how to modify the order of charge first/discharge first/ step default in cycle layer parameter setting in BTSDA software . If you have more questions, please contact me with: lulu@newarebattery.com https://www.facebook.com/batterytester.neware
If you tested a batch of batteries and want to compare the plots/curves of them, Neware BTSDA makes it possible easily, please follow the instruction here. Open BTSDA and choose the Plot/curve compare icon button. This is what the plot/curve compare function looks like, click the + sign to open 2 or more results in a bulk/batch. A fileopen dialog
Single record/data or multi records/data copy function of Neware BTSDA. It’s easy for you to copy single record/data or multi records/data out of Neware NDA data instead of output all of it. You can choose only one piece of data or use CTRL key to choose many pieces of data as you want. You can tell this from the above screen-shot.
A question about BTSDA comes from one of our customers. I am using BTS TestControl 5.3.0013 (2010.6.2) with an 8 channel battery tester, model BTS-3008-5V10mA.When I run a cell, it collects the data and shows every cycle, but after the data collection completes, the first eight cycles are missing (the data file starts with cycle #9).How can I prevent this
For some BTS3000 COM port edition battery testers customers, they may meet COM port errors sometimes. And for customers who are using TC5.3 software, when they check the testing data, there’s also possible for them to meet some minor problems in the BTSDA software. Let take a closer look on the following pictures: From the picture we can tell:On my
Generally, the capacity comes out from a battery(discharge capacity) will be less than the capacity comes into a battery(charge capacity). But for some customers, when they perform charge and discharge operation with Neware BTS software at the beginning, it’s strange to see the discharge capacity is more than the charge capacity. How’s that? There are 2 possible reasons. 1. It happens in