NEWARE Pack Tester Specification

CE-6004n-300V150A-IG

System Information
ParameterData
Channels4
Channel Parallel ConnectionSupport max 4 channels parallel mode
Input Power Supply3PH-AC380V ±10%, 50/60 ±5Hz
Power Factor≥99% (full load)
THDi≤5% (full load)
Input Power2 cabinets, each 100 kW
Input Current2 cabinets, each 151.9 A
Overall System Efficiency (Max)94% (regenerative)
Noise≤75 dB
Size (W×D×H)Each 1250 × 1050 × 1800 mm (2 cabinets)
WeightEach 1300 kg (2 cabinets)
Power Module TypeIGBT
Input Power WiringThree-phase four-wire system
Input ProtectionAnti-surge, anti-islanding, over/under-frequency, over/under-voltage, open-phase
Ingress ProtectionIP20
Measurement & Output
ParameterData
Voltage/Current SamplingFour-wire (same port for charging & discharging)
Voltage RangeCharge 0–300 V; Discharge 5–300 V (min discharge 5 V)
Voltage Accuracy±0.02% of FS; 24-bit resolution
Current Range0.75–150 A
Current Accuracy0.05% of FS (independent range); CV cut-off 0.05% of FS; 24-bit
Single-Channel Output Power45 kW
Total Output Power180 kW (4 channels)
Current Response Time (10%FS→90%FS)≤5 ms
Current Conversion Time (−90%FS→90%FS)≤10 ms
Minimum Step Time0.1 s
Input Impedance≥1 MΩ
Modes & Functions
ParameterData
Charge/Discharge ModesCCC/CCD, CVC/CVD, CC-CVC/CC-CVD, CPC/CPD, CP-CVC/CP-CVD, CV-CRD, CRD, VSS, CSS
Cut-off ConditionsVoltage, Current, ΔTime, Capacity, −ΔV
SimulationCurrent, Power; continuous charge/discharge switching; cut-off: Time, step line; up to 1,000,000 step lines
PulseCurrent, Power; min pulse 100 ms; up to 32 pulses; charge/discharge switch; cut-off: Voltage, ΔTime
DCIRBy calculation
Safety ProtectionSoftware: set safety limits (V/I min-max, delay…); Hardware: anti-reverse, OVP/OCP/OTP
Data & Analysis
ParameterData
Step Setting MethodForm editing
Recording Frequency100 Hz (10 Hz when two AUX channels allocate under one main channel)
DatabaseMySQL
Data OutputExcel, TXT (curve templates; customization supported)
Loop Test Range1–65535
Steps per Loop≤255
Loop Nesting≤10
Communication
ParameterData
Host CommunicationTCP/IP
InterfaceEthernet
Lower Computer Bandwidth1 M
Host Baud (Adaptive)10–100 M
Networking ModeLAN via switches/routers
Communication Expansion (Optional)CAN, RS485, BMS (DBC); supports integration with environmental chamber, water chiller, pressure fixture
AUX (Optional)
ParameterData
Temperature ChannelType-T: −70 to 260 °C; accuracy ±1 °C; resolution 0.1 °C
Voltage Channel0–5 V; accuracy ±0.05% of FS; resolution 0.1 mV
UsageBind with main V/I data; measured temperature can be used for control/protection
Environmental Requirements
ParameterData
Operating Temperature−10 to 40 °C (accuracy guaranteed within 25 ± 5 °C; drift 0.005% of FS/°C)
Storage Temperature−20 to 50 °C
Operating Humidity≤70% RH (no condensation)
Storage Humidity≤80% RH (no condensation)
Altitude<2000 m (above 2000 m: derate; every +100 m → −1 °C operating temp)